SYNCHRONOUSLY PUMPED OPO SPECIFICATIONS 1) |
Tuning range |
Signal |
1405 – 2020 nm (7115 – 4950 cm-1) |
Idler |
2250 – 4400 nm (4444 – 2253 cm-1) |
XIR |
5000 – 16000 nm (2000 – 625 cm-1) |
Linewidth |
< 5 cm-1 |
Output power 2) |
OPO @ 4000 nm (2500 cm-1) |
> 100 mW @ 2200 – 4000 nm |
OPO @ 12500 nm (800 cm-1) |
> 10 mW |
Pulse repetition rate |
87 MHz |
Pulse duration (pump laser) |
~8 ps |
Scanning step |
Signal |
0.1 nm |
Idler, XIR |
1 nm |
Polarization |
vertical |
Typical beam diameter 3) 4) |
~3 mm |
Typical beam divergence 5) |
< 5 mrad (for signal) |
Beam pointing stability |
< 50 µrad rms @ 1596 nm |
AOM modulation |
0 Hz – 2 MHz |
Fast spectral scan, scan speed (for spectral range) |
Idler |
< 2 s |
XIR |
< 1 s |
PHYSICAL CHARACTERISTICS |
Unit size (W×L×H) |
320 × 766 × 241 mm |
Power supply size (W×L×H) |
483 × 140 × 390 mm |
Umbilical length |
2.5 m |
OPERATING REQUIREMENTS |
Cooling |
water-air (by provided chiller) |
Room temperature |
22 ± 2 °C |
Room temperature stability |
± 1 °C |
Relative humidity |
20 – 80 % (noncondensing) |
Power requirements |
100 – 240 VAC (-10% / +5%), single phase, 50/60 Hz |
Power consumption |
< 1 kVA |
Cleanness of the room |
not worse than ISO Class 9 |